Protein Recovery Trial Sheet
Use this worksheet with a defined ECPVA configuration and your process records. Example inputs illustrate arithmetic only; replace them with your measured values.
Enter the operating or trial record
Protein Recovery Trial Sheet
Accepted-mass recovery (%) =100×accepted concentration(mg/L)×accepted volume(L) ÷[feed concentration(mg/L)×feed volume(L)].
Use representative mixed feed/accepted pools, or enter volume-weighted mean concentrations calculated from separately measured segment masses. A terminal grab sample cannot represent a whole campaign. Feed mass must be positive. Zero accepted mass is valid; >100% is an anomaly to investigate, not a capped result. Below detection limit is a bound, not exact zero. Recovery is product yield, not a direct adsorption measurement or ECPVA guarantee. Fields are user trial inputs; no supplier performance presets.
Use representative mixed feed/accepted pools, or enter volume-weighted mean concentrations calculated from separately measured segment masses. A terminal grab sample cannot represent a whole campaign. Feed mass must be positive. Zero accepted mass is valid; >100% is an anomaly to investigate, not a capped result. Below detection limit is a bound, not exact zero. Recovery is product yield, not a direct adsorption measurement or ECPVA guarantee. Fields are user trial inputs; no supplier performance presets.